High Cup Angle and Microseparation Increase the Wear of Hip Surface Replacements
Ian J. Leslie PhD, Sophie Williams PhD, Graham Isaac PhD, Eileen Ingham PhD, John Fisher PhD
Symposium: Papers Presented at the Annual Closed Meeting of the International Hip Society
Volume 467,
Issue
9
/
September ,
2009
Abstract on SpringerLink |
Full article HTML |
Full article PDF
Download Citation
Abstract
High wear rates and high patient ion levels have been associated with high (> 55°) cup inclination angles for metal-on-metal surface replacements. Wear rates and patterns have been simulated for ceramic-on-ceramic bearings by applying microseparation to replicate head offset deficiency. We tested 39-mm metal-on-metal surface replacements (n = 5) in a hip simulator with (A) an increased cup inclination angle of 60° and (B) an increased cup inclination angle and microseparation over 2 million cycles. (A) resulted in a ninefold increase in wear rate and (B) resulted in a 17-fold increase in wear rate compared to a standard gait condition study. Wear particles produced under microseparation conditions were larger than those produced under standard conditions but of similar shape (round to oval). The data suggest both head and cup position influence the wear of surface replacements; we believe it likely bearings with high wear either have a high cup inclination angle, an offset deficient head, or a combination of both.